CPAP Battery Backup Calculator

Pick your machine, the humidifier setting and your power source. See about how many nights it lasts on the AC outlet and on the maker's DC cable, and what size to buy for a medical device.

How we make money: some links on this page go to stores that may pay us a commission if you buy. It doesn't change the price you pay or which product we recommend; we rank by fit first. Details

Your machine

The heater uses most of the power. Ask your provider before turning it off.

Your outage, or until you can recharge.

Our figure for CPAP (heated humidifier and tube on): typical 56 W, range 40–75 W. Pressure and room humidity change it.

Your power source

1,024 Wh, 1,500 W continuous, UPS mode (switches over in 20 ms).

On Jackery Explorer 1000 v2, using the maker's DC cable, this lastsabout 1.9 nightsCould be 1.4 to 2.7 nights, depending on your machine and pressure setting.An estimate, not a promise.

That's short of your 3 nights. For a medical device we suggest 4,000 Wh or more.

AC outletDC cable
Energy a night618 Wh498 Wh
Nightsabout 1.5 nightsabout 1.9 nights
Size to buy4,900 Wh4,000 Wh

The DC cable skips the inverter and its idle draw, so it lasts 24% longer here. Plug it into the station's 12 V car port, and ask the machine's maker which cable fits. "Size to buy" covers twice your 3 nights with a 25% margin.

With the humidifier and heated tube off (10–30 W): about 7.2 nights. Check with your provider first; dry air bothers some people.

  • Medical device: these are estimates, not promises. Check with your provider and the device maker, and keep a non-electric plan.
Show the AC math
  1. Energy a night: (56 W ÷ 0.9 + 15 W idle) × 8 h = 617.8 Wh
  2. Usable capacity: 1,024 Wh × 0.95 = 973 Wh
  3. Nights: 973 Wh ÷ 617.8 Wh = 1.57
  4. Capacity to look for: 3 nights × 2 (medical) × 617.8 Wh × 1.25 margin ÷ 0.95 usable = 4,877 Wh

Assumptions used: Usable share of a power station; Medical planning multiplier; Planning margin; Inverter efficiency; Inverter idle draw; Inverter idle draw, scaled to size. What these mean

Show the DC math
  1. Energy a night: (56 W ÷ 0.9) × 8 h = 497.8 Wh
  2. Usable capacity: 1,024 Wh × 0.95 = 973 Wh
  3. Nights: 973 Wh ÷ 497.8 Wh = 1.95
  4. Capacity to look for: 3 nights × 2 (medical) × 497.8 Wh × 1.25 margin ÷ 0.95 usable = 3,930 Wh

Assumptions used: Usable share of a power station; Medical planning multiplier; Planning margin; DC port efficiency. What these mean

Power stations that meet the target (3 nights, the maker's DC cable)

  • EcoFlow DELTA Pro 3

    4,096 Wh · about 7.8 nights · UPS mode

    About $2,799Check price

  • Jackery Explorer 5000 Plus

    5,040 Wh · about 9.6 nights · UPS mode

    About $3,499Check price

  • EcoFlow DELTA 2 Max + 1 × EcoFlow DELTA 2 Max Smart Extra Battery

    4,096 Wh · about 7.8 nights

    About $1,928Check price

Each one holds twice your nights plus a 25% margin. Ranked by size, then price per usable Wh, with UPS mode as a tie-breaker; which stores pay us never affects the order. How we make money

How this is calculated

Energy a night

On the AC outlet, the machine's watts go through the inverter, which delivers about 90% of what it takes from the battery, and the inverter burns a few watts just by being on (the larger of 5 W or 1% of its continuous rating). For a CPAP that idle draw can be a third or more of the total. On the maker's DC cable there is no inverter: we only count a 10% loss in the cable or converter. We multiply by your hours a night and assume the station or inverter is switched off during the day.

Usable capacity

We count on 95% of a power station's rated Wh. For a 12 V battery we use 90% of a LiFePO4 battery and 80% of an AGM battery (fine in an emergency, though deep cycles shorten a lead-acid battery's life). Lead-acid also loses capacity at high current, which only matters for small batteries.

Nights and the range

Nights = usable Wh ÷ Wh a night, rounded down to 0.1. The range uses the low and high draw for each machine setting, since pressure, humidity and the machine itself all change the draw.

The size we suggest

For a medical device we plan for 2× the nights you ask for, then add a 25% margin: rated Wh = nights × 2 × Wh a night × 1.25 ÷ usable fraction, rounded up to the next 100 Wh. Outages run longer than forecast, batteries age, and a cold room or a higher pressure setting raises the draw.

What we don't know yet

Our CPAP-without-humidifier and BiPAP figures are estimates from typical ratings, not measurements, and inverter idle draw hasn't been measured on our own meter yet. If you know your machine's average watts (from its maker or a plug-in meter), pick "I know my machine's watts" for a better answer.

Full formulas and default assumptions: How the math works.